Coroner’s reportPrevention of future deaths · 2016-0112

“The Philips Respironics Trilogy 202 (BIPAP) (i)The BIPAP Trilogy 202 machine had a subtle…”

Dated 17 Mar 2016, from the coroner for Inner West London. Sent to 3 organisations or people.

A coroner wrote this report after an inquest, to warn the organisations below about a risk of further deaths.

Read the full report on judiciary.uk

How this works

The 56-day response deadline

Deadline passed

Replies were due by 12 May 2016. A reply may have been sent and not yet published.

Organisations that receive a report must reply to the coroner within 56 days, saying what they will do or why they will do nothing. We show whether a response is published on judiciary.uk.

  1. Report dated17 Mar 2016
  2. Deadline12 May 2016
  3. Published on judiciary.uk17 Mar 2016
  4. Today23 Oct 2026

Watch this report

We’ll email you when a response is published on judiciary.uk, or when the 56 days to reply end with none published there yet. Free, and you can stop with one click.

Next we ask your name, then send a link to confirm and create your free account. Already have an account? Sign in

Responses from each recipient

Responses from each recipient
Sent toResponse
St George's University Hospitals NHS Foundation TrustNamed in the report as “St Georges University Hospitals NHS Foundation Trust”Response publishedPublished on judiciary.uk in March 2016.
Phillips HealthcareDeadline passedNo response published on judiciary.uk so far. The 56-day deadline was 12 May 2016. A reply may have been sent and not yet published.
Department of Health and Social CareNamed in the report as “Department of Health”Response publishedPublished on judiciary.uk in March 2016.

Some responses were published as scans, so their text was read automatically. Check the wording against the original before you quote it.

The coroner’s concerns

This report was published as a scan, so its text was read automatically. Check the wording against the original before you quote it.

— (1) The Philips Respironics Trilogy 202 (BIPAP) (i)The BIPAP Trilogy 202 machine had a subtle visual display symbol which denoted when the machine was running on battery power. That symbol is not visible if there are many alert alarms as the alarm messages fill up the screen as they come in pushing the earlier alerts (including low battery) off the screen. In Mrs Scott’s case there had been 17 alert alarms in the space of 50 minutes. (ii) The BIPAP Trilogy 202 Machine has the same alarm sound for battery depletion as for circuit disconnect (where for example the face mask slipped) which was the more usual and expected reason for an alarm and these two factors separately and together did not have any feature of urgent warning to alert staff to battery depletion. (2) Staff training (iii) Staff who were experienced and trained on the BIPAP machine did not appear to be trained to be alert to the situation or to the significance of a battery symbol showing on the machine when the machine was plugged in to the mains or to any particular alarm which denoted battery depletion rather than mask slippage. (3) The provision of power to the ward (iv) Richmond ward ADU beds was designated as a category 4 area which in this case meant there was no isolated power supply (IPS) provided to the ward notwithstanding life-saving equipment was routinely used. (v) Hospital Technical Memoranda (HTM) 06-01 Part A provides advice and guidance and a benchmark standard for electrical installation, maintenance and safety etc in healthcare premises. It is a matter of concern that there is a conflict of advice between Clause 4.22 and Clause 6.62. Clause 4.22 states: “Clinical treatment and patient safety may be compromised ( but not endangered) by any interruption of electrical supply “ whereas Clause 6.62 states: “/n clinical risk Category 4 and 5 areas the patient environment should have at least two IPS circuits at the bedhead” (vi)There was no system or check that would alert ward staff to the failure of mains power in any particular area. (4) Notification to estates management (vii) The crash bell for bed bay 5 did not work when the emergency arose. However estates management had been notified some days earlier of the broken patient call in the same bay. This was of concern as both emergency bells were on the same circuit and not fixed until 2 April 2014 when by chance the failure of electricity was identified.

An extract. Read the full report, including the circumstances of the death, on the Courts and Tribunals Judiciary website.

Source
judiciary.uk, Prevention of Future Deaths reports, 2016-0112
Added to Minute Book
8 Oct 2026
Reuse
We show short extracts and link to the original. We don’t host the full report.

Something wrong, such as a date or who it was sent to? Report a mistake